Select a Diaphragm Seal or Chemical Seal | Forberg Scientific

Select a Diaphragm Seal or Chemical Seal

//Select a Diaphragm Seal or Chemical Seal

Select a Diaphragm Seal or Chemical Seal

Selection Guidelines
When selecting a diaphragm seal assembly, the following details must be taken into consideration to ensure a safe
and satisfactory operation. For specific technical assistance regarding temperature effects, volumetric compatibility, etc.,
contact the 
Forberg Scientific customer service department or send a completed diaphragm seal specification sheet to the factory
for analysis.

  1. Process Composition
  2. Temperature
  3. Pressure Range
  4. Pressure Instrument
  5. Process Connection
  6. System Fill Fluid
  7. Mounting Positions
  8. Response Time
  9. Seal and Gauge Matches

1. Process composition
Since the diaphragm and lower housing of the diaphragm seal will be exposed to the process medium, it is critical to
select materials for these components which will be compatible with this medium. Tables are available to assist in the
selection of these materials (see Pressure Gauge Section); however, the customer is the ultimate source for specifying suitable
materials.
WIKA cannot guarantee suitability. For information, see numerous reference guides such as corrosion table
reference books. If the pressure fluid is very thick, solidifies, or is full of solids, this should also be taken into consideration.

2. Temperature
Each diaphragm seal measurement system (diaphragm seal, pressure instrument, and cooling element or capillary, if
applicable) is filled with an amount of fill fluid at an ambient temperature of about 70oF. This temperature is referred to as
the system fill temperature. The fill fluid will expand or contract according to temperature changes. This in turn causes
the pressure in the sensing element to rise or fall, thus adding zero shifting effects to the instrument output. To reduce
this effect, the temperatures of the process and the environment should be specified when selecting a
diaphragm seal
system (see Diaphragm Seal Specification Sheet). Special advanced calibration techniques can be used to ensure the
best possible accuracy. At temperatures above 300 F, a cooling element or capillary is suggested to protect the pressure
instrument.

3. Pressure range
The displacement volume on the diaphragm seal required to “drive” each diaphragm seal measurement system
(
diaphragm seal, pressure instrument and capillary, if applicable) must be greater than the displacement volume needed
to move the pressure sensing element. Normally, the lower the pressure range, the larger the
diaphragm is required to
“drive” the system. Conversely, for higher pressure ranges,
smaller diaphragms are sufficient. Pressure transmitters also
follow the general rule of the lower the pressure, the larger the
diaphragm required.

4. Pressure instrument
As mentioned above (Item 3 – Pressure range), the diaphragm seal must supply sufficient displacement volume to
enable the
pressure instrument to reach full scale. As a general rule, smaller size gauges are better suited to low pressure
applications since less displacement volume is required on the part of the
diaphragm seal to drive the pressure instrument.

5. Process connection
The process connection is specified by the customer. Most process connections are threaded, flanged, or clamped;
however, additional connections are available. Teflon® coating and lining is only available in flanged connections, since
tapered NPT threads strip off the Teflon® during installation. However, solid Teflon® threaded connections are available
with NPT threads.

6. System fill fluid
WIKA offers a wide range of system filling fluids allowing temperatures from -130 F to 752 F. Chemical compatibility of the
system fill fluid with the process fluid must be carefully considered in the event of a leak. In food processing applications a
nontoxic fluid should be selected. Special fill fluids are also available for oxidizing media such as oxygen and chlorine.

7. Mounting position
Mounting position is important for diaphragm seal systems which include a capillary. The level difference between the
diaphragm seal and the pressure instrument causes a hydrostatic pressure to act on the sensing element:
a. For
gauges mounted above the level of the diaphragm seal, the pointer on the dial
of the
gauge will be lower than the zero point.
b. For
gauges mounted below the level of the diaphragm seal, the pointer on the dial
of the
gauge will be higher than the zero point.
The
diaphragm seal system can be calibrated to compensate for the effect caused by the hydrostatic pressure,
if the level difference is known in advance.

8. Response time
Response time, i.e., the time it takes the pressure instrument to indicate 90% of the value of a sudden pressure
variation, is especially important for instrument/
diaphragm seal assemblies which include a capillary. Response time
increases significantly in systems with long capillaries. In applications requiring long capillaries, response times can be
reduced by using larger diameter capillary tubing and reducing the viscosity of the system fill fluid. Be advised that
increasing the inner diameter of the capillary increases the temperature influence of the measuring system.
Forberg Scientific can consult WIKA if detailed information is needed.

9. Seal and gauge matches
For low ranges, gauge preference is 2XX.54 or 2XX.34 for access to perform calibration adjustments. Gauges with crimp
rings might not be usable due to potential recalibration. The table below shows the common matches between
gauge
and
diaphragm seal types recommended by the WIKA.

If you would like additional information about
WIKA Diaphragm Seals
or other
WIKA Products please
contact Forberg Scientific, Inc.